Short answer: ISO 9513 sorts extensometers into classes (0.2, 0.5, 1 and 2) according to how large their errors are when checked against a calibrator. A lower class number means tighter limits. Class 1 allows a bias error of ±1% of the reading or ±3 µm, whichever is greater. Class 0.5 halves that, to ±0.5% or ±1.5 µm. For a tensile test to ISO 6892-1, proof strength (Rp0.2) and yield strength need a class 1 or better extensometer. Modulus of elasticity measurement usually needs a better class, because the extensions involved are tiny.

What ISO 9513 actually checks
ISO 9513 is the calibration standard for extensometer systems used in uniaxial testing. During verification, a calibration laboratory compares the extensometer's output with a calibrator that displaces it by known amounts. The class awarded depends on several things together:
- Resolution of the reading.
- Bias (systematic) error of the extension reading, expressed as both a relative and an absolute limit.
- Gauge length error: how close the real gauge length is to the nominal one.
The class applies over a stated range and gauge length. An extensometer can be class 0.5 over part of its travel and class 1 over the rest, and the certificate should say so.
Class 0.5 vs class 1 in microns
| Class | Bias error limit (relative) | Bias error limit (absolute) |
|---|---|---|
| 0.5 | ±0.5% of reading | ±1.5 µm |
| 1 | ±1% of reading | ±3 µm |
Whichever limit is greater applies. That detail is what matters in practice. At small extensions the absolute limit (in µm) governs, and at large extensions the percentage does.
Worked example: which limit governs at Rp0.2?
Take a steel specimen with a 50 mm extensometer gauge length and a proof strength around 400 MPa. With an elastic modulus of about 200 GPa, the elastic strain at that stress is about 0.2%. Add the 0.2% plastic offset and the extension at Rp0.2 is roughly 0.4% of 50 mm, or 200 µm.
- Class 1: 1% of 200 µm is 2 µm, which is less than 3 µm, so the 3 µm absolute limit governs.
- Class 0.5: 0.5% of 200 µm is 1 µm, so the 1.5 µm limit governs.
Either error is small next to the 100 µm plastic offset, which is why class 1 is accepted for proof strength.
Why modulus is harder
Modulus is calculated from the slope of the elastic part of the curve, often over a strain window such as 0.05% to 0.25%. On a 50 mm gauge length, that is only 25 µm to 125 µm of extension. A 3 µm error is about 12% of a 25 µm reading. The same 3 µm is 2.4% of 125 µm. Errors of that size distort the slope, so labs that report modulus normally specify a better class, a longer gauge length, or both, and they average both sides of the specimen. Check the requirement in the standard you test to, as well as the customer's specification.
What ISO 6892-1 needs
ISO 6892-1 (IS 1608 Part 1 in India) calls for an extensometer of class 1 or better for proof strength and for upper and lower yield strength. For properties measured at larger extensions, such as percentage elongation at maximum force, it allows class 2. If you report all properties from one test, the extensometer has to meet the strictest requirement over the range used. ASTM E8 references the ASTM classification in ASTM E83, which uses different class names, so check which system your customer's specification uses.
Choosing an extensometer class
- Routine rebar, wire and plate tests for Rp0.2, Rm and elongation: class 1 is the usual requirement.
- Modulus, thin sheet, aerospace or R&D work: consider class 0.5, and confirm the gauge length suits the specimen.
- Large extensions, such as polymers or elongation to fracture: check the travel first. A class only means something within the calibrated range.
Akuracy offers electronic and mechanical extensometers, plus a non-contact video extensometer listed for a maximum of 800 mm extension. See the extensometer range. When you request a quotation, state the ISO 9513 class, gauge length and travel you need, so the supplied unit and its calibration certificate match your test standard.
Reading the calibration certificate
Check that the certificate states the class, the gauge length, the range over which the class applies, the calibrator's traceability and the date. A certificate that only says "class 1", without a range, can't show whether the class covers your Rp0.2 region.
FAQ
What does class 0.5 mean for an extensometer?
It means that under ISO 9513 its errors, including a bias error of ±0.5% of reading or ±1.5 µm (whichever is greater), stay inside the class 0.5 limits over the calibrated range.
Is class 1 enough for a tensile test?
For proof strength and yield strength under ISO 6892-1, yes. Class 1 or better is required.
Which class do I need for Young's modulus?
Usually better than class 1, because the elastic extensions are only tens of microns. Check your test standard and the customer's requirement.
Can one extensometer have two classes?
Yes. A certificate can show different classes over different ranges of travel.
What is the ASTM equivalent of ISO 9513?
ASTM E83, which uses its own class designations.
Request a Techno-Commercial Quotation
Share your materials, specimen sizes, gauge lengths and the class your standard requires, and we will suggest a matching extensometer and UTM setup.